|
July 05
Flash
Steam Sterilization, Quality Assurance and
the Operating Room: A Practical
Guide
by
Arthur Henderson, RN; Gaylyne Marshall,
RN; clinical education specialists; Susan
Fagan, RN, senior clinical education specialist,
STERIS Corporation
Quality control of sterilization processes
is critical for reducing the chance of nosocomial
infections and improving patient safety.
This is especially true when items are flash-sterilized
in the operating room (OR) and immediately
utilized for a patient. The immediacy of
use places an extra focus on the quality
control practices employed by the OR to
ensure that devices that are flash sterilized
are safe at the time of use. Due to the
conflicting recommendations of biological
and chemical indicator manufacturers, the
quality control protocols for flash sterilization
processes vary greatly throughout the medical
industry. This can leave any OR manager
wondering, "What must I do to ensure
that the flash sterilization processes used
in my OR meet best practices standards?"
The surest way to maintain best practices
is to be thorough in your quality assurance
process.
Quality control of the flash sterilization
process should include three critical functions:
1. Equipment qualification
2. Product testing
3. Routine monitoring
Equipment Qualification
Sterilization quality control in the OR
starts with the equipment. Without properly
functioning equipment, sterilization is
impossible. Equipment qualification must
be performed when the sterilizer is first
installed and after any major repairs. Two
tests are used in this qualification: biological
testing for all cycles and air removal qualification
testing for prevacuum cycles only. In addition
to these tests, the electrical supply water
supply and other mechanical requirements
for the unit must be verified whenever systems
are installed or moved.
Biological Qualification
The biological qualification involves performing
three consecutive biological challenges
for each cycle type and flash sterilization
configuration. First, generate a list of
all the flash sterilization cycles that
are or will be used on the sterilizer. This
list would include gravity flash at 270‰F,
prevacuum flash at 270‰F, and any other
cycles you use. Next, list all the types
of configurations sterilized with these
cycles; open mesh trays, protective organizing
cases (rigid containers), rigid sterilization
containers appropriate for flash, etc. The
goal of this exercise is to identify all
the possible combinations of cycles and
configurations that would be used when flash
sterilizing items.
The last step is to perform three consecutive
biological test cycles with each test condition.
Always choose the shortest exposure time
if more than one exposure time is used for
a given cycle and configuration.
If you are flash-sterilizing an open mesh
tray for both 3 minutes and 10 minutes,
only the 3 minute exposure time is tested.
A minimum of one biological indicator is
placed within each configuration. It is
very important that no items are placed
in the trays, cases, or containers. The
instrumentation would slow the come-up time
of the steam sterilizer an allow more kill
during the come-up time. A successful qualification
occurs when all the biological indicators
demonstrate no growth. For more information
on the specifics of the testing process,
refer to ANSI/AAMI ST37.
Several choices of biological indicators
exist today for this application. The most
common are self-contained biological indicators
(SCBIs). SCBI contain bacterial spores impregnated
into a piece of filter paper and housed
within a vial containing sterile growth
medium. The bacterial spores are cultured
and incubated after the sterilization cycle
by breaking the ampoule of growth medium
within the vial and incubating the vial
as directed by the manufacturer, typically
24 to 48 hours. A second type of SCBI uses
an enzyme-based early-readout component
of one or three hours, with traditional
spore growth assay of 24-48 hours. It must
be noted that for this critical test, the
Association for the Advancement of Medical
Instrumentation (AAMI) recommends conventional
BI outgrowth. In the section entitled "Biological
indicators with enzyme-based early-readout
capability," AAMI states that "early
read-out capabilities should not be relied
upon for these critical assessments,"
which includes equipment qualification upon
installation and after major repairs.
Air Removal Qualification
The air removal qualification is only required
for those steam sterilizers with a prevacuum
sterilization cycle. The qualification consists
of three consecutive air removal tests (Bowie-Dick
tests) to demonstrate that sufficient air
can be removed by the sterilizer from the
representative porous item. Many sterilizer
manufacturers have preprogrammed test cycles
for this test. Detailed instructions for
the construction and testing of Bowie Dick
Test Packs can be found in the AAMI standard
ANSI/AAMI ST46. A successful qualification
occurs when all the Air Removal Test Cycles
demonstrate complete air removal.
Equipment qualification is the first required
element of a good quality control program.
Failure of any part of this qualification
indicates that the sterilizer is malfunctioning
and requires servicing or repair. When servicing
is complete, all qualification testing must
be successfully completed before the sterilizer
can be put back into service.
Product Testing
Product testing is the next step in a good
quality control process.
This testing uses biological indicators
to confirm that the cycles and configurations
used within the facility for flash sterilization
effectively kill microorganisms on the items.
Product testing should be performed periodically
(once a year) and any time there is a significant
change to the process. Some examples of
significant changes include:
Change in the exposure time or temperature
Change in the cycle type (gravity
to prevacuum)
Change in the packaging configuration
(wrapped trays to a rigid sterilization
container)
Receipt of a new type of instrument
To perform product testing, start with the
same list of cycle types and configurations
identified during the qualification of the
sterilizer, but this time add to the list
the type of instruments and quantity being
flash-sterilized in a single cycle. Next,
test the easiest and hardest combinations
to sterilize for each cycle type and configuration.
Typically the easiest is the load with the
smallest and fewest items. The hardest load
would most often be the largest one with
the most instruments, though exceptions
do exist. If you are doing product testing
for the first time, you should consult with
your sterile processing manager for help
with the testing.
Routine Monitoring
After the equipment has been qualified and
the procedures and processes of sterilization
have been verified, routine monitoring is
performed according to AAMI standards to
capture any shifts in the performance of
the equipment or personnel that would change
the effectiveness of the sterilization process.
Routine monitoring helps to ensure that
the equipment is still working and the operators
have not made a mistake.
Biological Monitoring of the Sterilizer
Equipment performance is verified weekly
using biological indicators. A single biological
indicator test is performed for each of
the cycle types and configurations that
are used. To perform the test, a biological
indicator is placed within an otherwise
empty configuration such as an open mesh
tray, protective organizing case (rigid
container), or rigid sterilization container
appropriate for flash. The item is processed
in an otherwise empty sterilizer chamber,
using the cycle type identified for the
configuration being tested and the shortest
exposure time. Several choices of biological
indicators are used for this application.
The most common are the traditional SCBIs
and those with the additional enzyme-based
component. AAMI recommends that an enzyme-based
early-readout biological indicator complete
its full incubation time to test for spore
growth on a weekly basis.
Air Removal Monitoring of the Sterilizer
Air removal testing (Bowie-Dick) is required
to be performed each day that a prevacuum
cycle is run. The test must be performed
before the prevacuum flash cycle is run.
Monitoring Flash Cycles
Each sterilization cycle must be monitored.
This is accomplished by reviewing cycle
printouts and chemical indicator strip results.
Every tray, container, or other configuration
must contain a chemical indicator during
sterilization. The chemical indicator is
placed in the most difficult-to-penetrate
location within the configuration. Several
types of chemical indicators are available
for monitoring the flash sterilization cycle.
AAMI has defined each type of chemical indicator
and its performance.
CLASS 1: The most basic indicator is the
process indicator, which includes such things
as tapes, data cards and indicator strips.
Process indicator strips simply verify that
the cycle completed, but are not necessarily
linked to the achievement of critical sterilization
parameters like time, temperature and steam
quality. Process indicators give you minimal
assurance that steam was present but not
that the parameters of sterilization were
met.
CLASS 4: Class 4 indicators provide more
sterility assurance than Class 1 indicators
by monitoring the achievement of 2 or more
critical parameters for sterilization (time,
temperature, or steam quality). Class 4
indicators are also held to a higher standard
of accuracy and allow a narrower tolerance
than process indicators. This means the
indicator change occurs much further into
the sterilization cycle.
CLASS 5: The highest level of assurance
is
provided by the Class 5 integrating indicators.
Class 5 indicators monitor all the critical
parameters of steam sterilization (time,
temperature and steam quality) and have
been correlated to the death curve of a
biological indicator. This means that the
Class 5 integrating indicator will demonstrate
a "pass" when sterilization conditions
were met that would kill a biological indicator,
but will demonstrate a "fail"
if the conditions would not have killed
a biological indicator. Class 5 indicators
are also held to a higher standard of accuracy
and maintain even narrower tolerances than
Class 4 indicators. As a predictor of biological
response, Class 5 integrators provide a
high assurance level.
What about Implants?
The Association of peri-Operative Registered
Nurses (AORN), The Centers for Disease Control
and Prevention (CDC) and the Association
for the Advancement of Medical Instrumentation
(AAMI) all agree that implants should not
be flash-sterilized. The implant is placed
in an avascular environment that would promote
infections from improperly sterilized or
contaminated implants. Since flash sterilization
can increase the risk of contamination of
the implant during transport from the sterilizer
to the patient, flash sterilization of implants
should be avoided. The need to flash-sterilize
implants can be eliminated through careful
planning and inventory management.
Should a situation exist in which an implant
must be flash-sterilized, both AAMI and
AORN recommend monitoring the cycle with
a biological indicator. The implant must
then be quarantined until the result of
the biological testing is known. Of the
biological indicators available on the market,
the SCBIs with enzyme-based early-readout
capabilities give the fastest results in
one or three hours, depending on the product
used. However, even one hour is a long time
to expose an unpackaged flash-sterilized
implant to environmental contaminants.This
creates a dilemma for the facility: should
they quarantine the implant until the biological
results are known and increase the potential
for contamination of the unwrapped implant,
or should they use the implant without waiting
for the biological result?
One solution proposed through AAMI is a
procedure for releasing the sterilized implants
prior to knowledge of the biological indicator
results. In this proposed method, the implant
is released using other means to ensure
that the sterilizers performance was
adequate. Cycle printouts and other indicators
such as the Class 5 indicators are used.
This method requires documentation of the
early release and subsequent follow-up with
the biological indicators final results,
but it can be used to reduce the risk of
environmental contamination of a flash-sterilized
implant before it is implanted in the patient.
Ultimately, the decision to flash-sterilize
an implant is the facilitys responsibility.
The decision should be based on several
considerations, including the hospitals
surgical site infection rates, the location
and configuration of the surgical suites
in relationship to the sterilizer, and the
assessment of a patient's risk for a surgical
site infection. The decision to flash-sterilize
an implant should never be based upon the
need to decrease instrument turnaround time,
an inventory shortage of surgical instruments,
or the desire to save time.
Conclusion
The success of any quality control protocol
for flash sterilization rests on the practices
employed by the facility's central service
and surgical professionals. Policies and
procedures are needed to provide instruction
and guidance and to document the total quality
control process from the cleaning and preparation
of the instrument through the use of that
instrument on a patient. It is also imperative
to reinforce those procedures by providing
ongoing training to anyone who is operating
the sterilizers. Given a choice, terminal
sterilization is the preferred method of
reprocessing instruments However, flash
sterilization can be performed successfully
and responsibly in the surgical suite by
establishing and following a thorough quality
control program.
ANSI/AAMI ST37 "Flash Sterilization:
Steam sterilization of patient care items
for immediate use", 1996
ANSI/AAMI ST46 "Steam sterilization
and sterility assurance in health care facilities",
2002
ANSI/AAMI ST46 "Steam sterilization
and sterility assurance in health care facilities",
2002
ANSI/AAMI ST60 "Sterilization
of healthcare
productschemical indicatorsPart
1: General Requirements"1996
AORN "Standards, Recommended
Practices, and Guidelines", 2004
The Centers for Disease Control and
Prevention (CDC) "Guidelines for Prevention
of Surgical Site Infection", 1999
|
|